The K450X Carbon Coater with Flash Evaporation
The K450X uses several Carbon Fibre types to cover a range of deposition thicknesses operating at rotary vacuum pressures. Coating is omni-directional, and
this fully Automatic System has a relatively short cycle time of the order of 5 minutes.
The system employs low voltage, high current, electrodes between which the carbon fibre is located.
The electrode assembly employs a protective shutter to protect the specimen from the heat radiation damage during the timed outgassing of the carbon fibre process. At completion of this process, the shutter automatically retracts, and full power is applied to the electrodes, causing the carbon to burn quickly or 'flash' the fibre fusing as part of the process. At this point, the shutter returns to the protective position.
The controls can be preselected to allow for different types of carbon fibre.
This, together with adjustable specimen stage to electrode distance, and fully automatic control, allows repeatable deposition thickness to be achieved. The vacuum cycle and external rotary vacuum pump, being fully controlled by the Instrument.
Features:
- Fully automatic control
- Automatic protection shutter
- Even thickness deposition over a range of coatings using string or cord
- Micro-controller fully automated with manual option system control with stored parameters
- Menu-driven 'user' key inputs
Benefits:
- Easy to operate
- Protects sample during out-gassing of the carbon fibre
- Reproducible coatings
- Easy operation for multiple users
Specifications of the K450X Carbon Coater:
Instrument Case - 450mm W x 350mm D x 175mm H
Work Chamber - Borosilicate Glass 165mm Dia x 125mm H
Safety Shield - Polycarbonate
Weight - 18Kg
Carbon Source - Carbon Fibre
Specimen Stage - 60mm Dia. Rotating Specimen Stage with Tilt Facility. (Spacing to Electrodes 60mm)
Vacuum Gauge Range - ATM - 1x10-4 mbar
Ammeter Gauge - 0.50 Amps
Low Voltage - 25 Volts
Outgas Current - Selectable for type of Carbon source
Services - Nitrogen as dry pure gas (if required) at nominal 5 psi
Vacuum Pump - Double Stage Rotary vacuum Pump complete with vacuum hose and Recommended Oil Mist Filter.
Electrical Supply - 230 Volts 50Hz (10 amp max. including Pump)
115 Volts 60Hz (16 amp max. including Pump)
Product Conformity